Autoclave beech kaam me kharab ho gaya — patients hain, register me kaise dikhau?

Machine down entry Compliance Logs/PMS me record karo, backup sterilization arrangement (doosra autoclave/pouch service) use karo, aur repair + re-validation (BI pass) ke baad hi wapas chalu maano.

Where in ORALZY: NABH Hub → Daily Logs → Compliance Logs / Registers (pms, capa)

Diagnosis: Autoclave down hone se sterile instruments ki supply ruk sakti hai. NABH me breakdown chhupana nahi, manage karna dekha jata hai.

Karo:

  1. NABH Hub → Daily Logs → Compliance Logs (ya Registers → 'pms'/'calibration') me breakdown ka time, kya hua, kisko bulaya — sab log karo.
  2. Backup: doosri machine, ya pre-sterilized disposable/pouched instruments use karo. Jab tak sterile supply na ho, elective invasive procedure tal do.
  3. Repair ke baad seedha kaam mat shuru karo — pehle BI (spore-test) + TST strip PASS karao, phir 'validated' maan ke chalu karo.
  4. NABH Hub → Registers → capa me note: down time, patient impact, preventive (AMC/PMS schedule).

Aage se: NABH Hub → Compliance Logs me equipment PMS/AMC schedule daalo taaki preventive maintenance overdue na ho. Ek backup sterilization plan SOP me likho.

Edge case: Agar backup na ho aur invasive kaam ho gaya — usse incident report banao (Incidents), warna infection risk documented nahi dikhega.

Quick steps

  1. Breakdown time + action Compliance Logs me record karo
  2. Backup sterilization (2nd machine/pouched) use karo
  3. Invasive elective procedure supply theek hone tak tal do
  4. Repair ke baad BI + TST pass karao, phir chalu maano
  5. capa me down-time + preventive (AMC/PMS) likho