IPC.5 pe assessor sterilization sabse zyada grill karta hai. Ratta answers: standard cycle 121°C @ 15 psi, 30 min (Class B autoclave dental ka gold standard; flash 134°C/30psi/4min). Instrument reprocessing 7-step chain — pre-soak → clean → rinse → dry → pack → sterilize → store dry (shelf-life ~21 din). Chemical indicator har cycle, Bowie-Dick (steam-penetration) roz vacuum autoclave pe, biological indicator/spore test HAFTE me ek baar + har repair ke baad + har implant load ke saath. Killer sawaal: 'Spore test POSITIVE aaye to?' — Jawab: last successful BI ke baad ki saari loads quarantine karo, re-process, engineer service, infection control ko batao, exposed patients ka recall list banao, CAPA karo. ORALZY me proof: Registers → Autoclave / Spore Test / TST-Bowie-Dick. READY rakho: last 3 mahine ki spore-test entries + Class B autoclave ka log.
NABH me assessor sterilization aur autoclave ka kya poochta hai?
Autoclave 121°C/15psi/30min (Class B), daily Bowie-Dick, weekly spore test (BI). BI positive aaye to last negative se saari loads quarantine + patient recall.